Chapter 3 Handling Plates
3.2 Plate end processing
When plate end is displayed, remove the plate and re-set a new one following
the procedure below.
1) When is displayed on the panel:
The plate ended during exposure, so it is discharged from the processor.
Stop the buzzer by pressing the (reset) key.
After the plate is discharged, press the key to shift to the local mode,
and mount a new plate following the procedure mentioned in "3.1 Plate
setting".
Whether or not data will be
exposed depends on the
amount of exposure data
remaining when the plate end
is detected.
2) When is displayed
on the panel:
Plate ended at the beginning of an
exposure, so the remainder is
shorter than the processor's
minimum transfer length. Please
remove the plate.
1. Open the front door and remove
the plate through the plate removal
port.
2. Close the door and press the
key to return to the local mode.
3. Re-set the plate following the
procedure mentioned in "3.1 Plate
setting".
Pick up plate
Note
MENU
<Plate End>
- 23 -
Cover for the plate
removal port
Plate removal port