Agilent Technologies E6385A Whiteboard Accessories User Manual


 
Reference
Parameters
Chapter 5
119
Reference
2. GN MSC Command Speed [0=slow 1=fast]
Enter the selection to determine the speed at which commands will be sent from the
Test Set through the PC to the MSC during automated testing.
If you select 0, the measurements will be made in the most reliable manner.
If you select 1, the measurements will be made in the most expeditious manner.
This setting is used most often when using a PC running the BTS Laptop Utility
and an internal modem.
3. GN Power Output [0=meas 1=spec 2=blnk]
Enter the selection to determine the manner in which output power test results will
be displayed.
If you select 0, the measured output power will be entered in the Output Power
column of the FCC short form when the Full Site Test (FCC) test or Single Radio
Test is run. Also, when the TX Power Adjustment test is run, the center scale
value on the meter will be determined by the pass/fail limits.
If you select 1, the value that you specify for the radio in the frequency plan, or at
the start of the Single Radio Test, will be printed in the Output Power column of
the FCC short form (not the measured value). The measured value will still be
printed in the long form. Also, when the TX Power Adjustment test is run, the
center scale value will be set to the value that you enter for the Specified
Output Power at the start of the test.
If you select 2, the Output Power column of the FCC short form will be left blank.
4. GN Stop Test if results fail [0=no 1=yes]
Enter the selection to determine the action of the Test Software in the event of a
measurement failure. An F will appear in the P/F column of the test results to
indicate a failure of that data point.
If you select 0, the Test Software will continue running if the measured value is
not within specifications.
If you select 1, the Test Software will stop testing if the measured value is not
within specifications.
5. INST EXT REF IN [1,2,5,10,15 MHz]
If using an external reference for the Test Set timebase, use this field to select the
appropriate reference frequency.
6. INST EXT REF SELECT [1=AUTO 2=INT 3=EXT]
This setting determines the source for the Test Set timebase:
Auto - Uses the Test Set internal timebase, unless an external reference signal of
sufficient amplitude and correct frequency is applied to the REF IN connector.
INT - Always uses the Test Set internal timebase.
EXT - Requires that an external timebase be connected to the REF IN connector.
You must also specify the external timebase signal frequency in Parameter 5 -
INST EXT REF IN [1,2,5,10,15 MHz].