Agilent Technologies E6385A Whiteboard Accessories User Manual


 
Troubleshooting
Errors When Loading and Running the Test Software
Chapter 6
O:\Manuals\E6385A_Amps\Book\Trouble.fm
132
Errors When Loading and Running the Test Software
This group of errors would most often be encountered before the Main Menu appears.
Such errors would be most likely to occur the first time that you load and run the Test
Software (see Chapter 2, “Installation,” on page 21). The errors are presented with the
error text, followed by a description of the cause, followed by some possible solutions.
Symptom: Memory Overflow Errors
The random access memory (RAM) space of the Test Set is shared by IBASIC
programs and save/recall registers. If you have saved a large number of registers
in the Test Set, a memory overflow error might occur when you first attempt to
load the Test Software.
To correct this, clear some RAM space by deleting some or all of the saved
registers. You may do this selectively (one register at a time) or globally (all
storage registers at once).
NOTE Clearing the registers, whether selectively or globally, is permanent. Registers are not
retrievable once deleted. You may save registers to an SRAM card, as well. Refer to
the Agilent Technologies 8935 Reference Guide for instructions on this procedure.
To delete a single register:
Step 1. Press the
Recall key.
Step 2. Scroll the cursor to the register to be deleted.
Step 3. Press the
Yes On/Off key to indicate that you wish to clear the register.
Step 4. Press the
Yes On/Off key again to confirm.
Step 5. The register will be cleared.
To delete
all
saved registers:
Step 1. Press the
Recall key.
Step 2. Scroll to *Clr All* and press the knob.
Step 3. Press the
Yes On/Off key to confirm.
Step 4. All save registers will be cleared.