Emerson Network Router Network Router User Manual


 
Installation and Use Manual 61
Transmitter Configuration, Characterization, and Calibration
Meter Verification Transmitter TerminalsData LoggerTransmitter Configuration
Figure 5-5 D1 and D2 density calibration
Fill sensor with D1 fluid
Calibration
Density cal - Point 1
Do Cal
Enter density of D1 fluid
Fill sensor with D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density cal - Point 2
Do Cal
Enter density of D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Close Close