62 ProLink
®
II Software for Micro Motion
®
Transmitters
Transmitter Configuration, Characterization, and Calibration
Figure 5-6 D3 or D3 and D4 density calibration
Fill sensor with D3 fluid
Calibration
Density cal - Point 3
Do Cal
Enter density of D3 fluid
Fill sensor with D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density cal - Point 4
Do Cal
Enter density of D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Done
Close Close