Lightwave Communications LDX-3232 Power Supply User Manual


 
COMMAND REFERENCE
LDX-3232 Device-Dependent Commands
04_06 LDX-3232 39
CHAPTER 4
LDX-3232 Device-Dependent Commands
This section contains all of the device-dependent commands for the LDX-3232
High Compliance Current Source, listed in alphabetical order. Sub-sections for
each path are presented, listing the commands that are legal for that path. See
Figure 3.1 for command path tree structure.
Figure 4.1 Command Description Format
Command Path:Name
Re
q
uired letters are shown in
u
pp
er case; lower case letters
are o
p
tional.
Front Panel/
A solid box means the command
can be executed in that mode.
Command mnemonics in front
p
anel
(
local
)
mode ma
y
var
y
.
Command Descri
p
tion
S
y
ntax Dia
g
ra
m
(
For Remote Commands
)
Parameters
Tells what
p
arameters are
ex
p
ected, and their t
yp
e.
An
<nrf value>
refers to a
flexible numeric notation
(
IEEE-488.2
)
. Inte
g
er,
ex
p
onential, or floatin
g
p
oint
ma
y
be used with the decimal
radix. Alternate radices ma
y
be
selected in remote o
p
eration.
Points of Interest
Has s
p
ecial information
about the command.
Exam
p
les
Shows t
yp
ical statements
using the command.
Front Panel
Remote
LASer:LIMit:I
limit value.
SYNTAX DIAGRAM
<white
space>
< nrf
value >
:
:
:
LAS
LIM
PARAMETERS
An <nrf value> that represents the LASER limit
current, in mA.
POINTS OF
INTEREST
The current limit is in effect in all modes of
operation of the laser output.
In local operation, the limit current is entered
by selecting the LIM I parameter, pressing and
holding in the (PARAMETER) SET switch,
adjusting the ADJUST knob until the desired value
is displayed, and then releasing the SET switch.
EXAMPLES
Remote Execution
I2
"Las:Lim:I2 100" - action: the LASER current limit
is set to 100.0 mA.
"LAS:LIM:I4 3000" - action: the LASER current limit
is set to 3000.0 mA.
1
The LASer:LIMit:I2 command sets the laser current