Lightwave Communications LDX-3232 Power Supply User Manual


 
COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
04_06 LDX-3232 63
CHAPTER 4
The LASer:LDV command sets the laser voltage for calibration of the laser voltage
measurement.
Syntax Diagram
Parameters An <nrf value> that represents the (laser) output voltage, in volts.
Notes setpoint is the same for both low and high bandwidth output modes and is entered only
for LDV calibration.
Examples "Las:LDV 4" -action: enters a value of 4.000 V.
"LAS:ldv 1.025" -action: enters a value of 1.025 V.
The LASer:LDV? query returns the value of the measured laser voltage.
Syntax Diagram
- where the response is the measured laser output voltage for either low or high
bandwidth modes.
Notes This measurement is updated approximately once every 600 ms.
Examples "LAS:ldv?" -response: 3.03, means the measured laser output voltage is 3.03 V.
"Laser:LDV?" -response: 1.0, means the measured laser output voltage is 1.000 V.
LASer:LDV
FRONT PANEL
REMOTE
LASer:LDV?
FRONT PANEL
REMOTE
<white
space>
<nrf
value>
:
:
LAS LDV
:
LAS
:
LDV
?