COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
04_06 LDX-3232 59
CHAPTER 4
.
The LASer:ENABle:OUTOFF? query returns the value of the status outoff enable register
of the LASER operations.
Syntax Diagram
-where the response is the sum of the following:
Notes The enabled LASER events can be set by using the LASer:ENABle:OUTOFF command.
The LASER output status can be monitored by the LASer:EVEnt? query.
Examples "LAS:ENAB:OUTOFF?" -response: 1, means that Current Limit Condition will cause the
LASER output to be turned off.
"Radix Hex; Las:Enab:Eve?" -response: #H20B, means that all of the above conditions
will cause the LASER output to be turned off.
LASer:ENABle:OUTOFF?
FRONT PANEL
REMOTE
1 LASER Current Limit 256 N/A
2 LASER Voltage Limit 512 Output if out of tolerance
4 N/A 1024 N/A
8 Power limit (with output on) 2048 N/A
16 N/A 4096 N/A
32 N/A 8192 N/A
64 N/A 16384 N/A
128 N/A 32768 N/A
:
LAS
:
ENAB
OUTOFF
:
?