COMMAND REFERENCE
LDX-3232 Device-Dependent Command Reference
44 LDX-3232
CHAPTER 4
The LASer:CAL:LDI? query is used to determine that the LDX-3232 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:LDI mode.
Syntax Diagram
Notes After this query is issued and a response of 1 is received, the LDX-3232 will be ready for
the user to enter a current value via the LASer:LDI command (see Chapter 6).
Examples "LASer:CAL:LDI?" -response: 1, means the LDX-3232 is ready for the user to enter a
current value via the LASer:LDI command.
“LASer:Cal:LdI?” -response: 0, means the LDX-3232 is not yet ready for the user to enter
a LASER current value
The LASer:CAL:LDV command is used to enter the LASER voltage measurement
calibration mode.
Syntax Diagram
Parameters None
Notes After this command is issued, the LDX-3232 will allow calibration of the laser voltage
measurement. This procedure is outlined in Chapter 6.
The LASer:CAL:LDV? query may be used to determine if the LDX-3232 is ready for the
user to enter a value.
Examples "Las:CAL:Ldv" -action: the LDX-3232 enters calibration mode for LASER voltage.
"Las:Cal:LDV" -action: the LDX-3232 enters calibration mode for LASER voltage.
LASer:CAL:LDI?
FRONT PANEL
REMOTE
LASer:CAL:LDV
FRONT PANEL
REMOTE
:
LAS
:
CAL
I
:
?
LDI
:
LAS
: :
CAL
LDV