Appendix B: Remote Connectors
1760 Series Component Waveform/Vector Monitor User Manual
B-4
Remote Connector Converter
If the 1760--Series i s used as a dire ct repl acement for a
1740--Serie s or 1750--Series instrument, construct an adapter
or rewire the remote cable according to Figure B-2. If the
1760--Serie s replaces an Option 16 instrument, construct the
adapter in the same manner, omitting the pin 3-to-pin-25
connection.
13
25
RGB INPUT
RGB ENABLE*
REMOTE SYNC IN
GROUND
REMOTE SYNC ENABLE
3
19
6
7
5
Remote Plug
14
1
1760--Series
Rear-Panel
REMOTE Connector
25
8
9
10
1
14
13
* (Do not connect when replacing Option 16 instruments.)
Figure B-2. Using t he 1760---Series as a 1740/1750---Series replacement.