Operator’s Checkout Procedure
1760 Series Component Waveform/Vector Monitor User Manual
2-12
Initial Equipment Connections
H Connect the 1760--Series to an appropriate AC power
source.
H Connect the generator NTSC output (COMPOSITE for
TSG131) to the CH-A input. Connect the CH-A loop-
through to the EXT REF input and connect a 75Ω
terminator to the EXT REF loop-through. See
Figure 2-4.
H Connect the generator Y output to the CH-A1 input.
Connect B--Y to CH-A2 and R--Y to CH-A3. Connect a
75Ω terminator to each loop-through.
H Set the generator FORMAT to NTSC (COMPOSITE for
PAL ).
A
A1A2
A3
BB1
B2
B3
EXT
REF
1760---SERIES
NTSC
NTSC
Y
Y
B-YR-Y
TSG 130---SERIES
Figure 2-4. Equipment hook-up for Operator’s Checkout Procedure.