Agilent Technologies E1442-90003 Switch User Manual


 
56 E1442A Command Reference Chapter 3
ARM
The ARM subsystem allows a scan list to be scanned multiple times
(1 through 32767) with one INITiate command.
Subsystem Syntax ARM
:COUNt <number> MIN | MAX
:COUNt? [MIN | MAX]
ARM:COUNt
ARM:COUNt <number> allows scanning cycles to occur a multiple of
times (1 to 32767) with one INITiate command and when
INITiate:CONTinuous OFF | 0 is set.
Parameters
Comments Number of Scans: Use only values between 1 (MIN) to 32767 (MAX) for
the number of scanning cycles (default is MIN = 1).
Related Commands: ABORt, INITiate:IMMediate,
INITiate:CONTinuous
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
$50&281
Sets 10 scanning cycles
6&$1#
Sets channel list
,1,7
Starts scanning cycle
ARM:COUNt?
ARM:COUNt? [MIN | MAX] returns the current number of scanning cycles
set by ARM:COUNt. If a value between MIN and MAX is set, that value
for ARM:COUNt is returned. The optional parameters MIN and MAX
allow you to query the module for the these values instead of looking
them up in the command reference. A 1 is returned for the MIN
parameter. 32767 is returned for the MAX parameter regardless of the
ARM:COUNt value set.
Name Type Range of Values Default Value
<number> numeric 1 - 32767 | MIN | MAX 1