Table C-2. E1442A Error Messages
Code Error Message Potential Cause(s)
-211 Trigger ignored Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
-213 Init Ignored Attempting to execute an INIT command when a scan is already in
progress.
-222 Data out of range Parameter value is outside valid range.
-224 Illegal parameter value Attempting to execute a command with a parameter not applicable to
the command.
-240 Hardware error Command failed due to hardware problem.
-310 System error, Internal driver
error.
This error can result if an excessively long parameter list is entered.
1500 External trigger source already
allocated
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.
1510 Trigger source non-existent Selected trigger source is not available on this platform (e.g., some
triggers are not available on VXI B-size mainframes).
2000 Invalid card number Addressing a module (card) in a switchbox that is not part of the
switchbox.
2001 Invalid channel number Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g.,, channel 99 of a multiplexer module).
2006 Command not supported on this
card
Sending a command to a module (card) in a switchbox that is
unsupported by the module.
2008 Scan list not initialized Executing a scan without the INIT command.
2009 Too many channels in channel
list
Attempting to address more channels than available in the switchbox.
2010 Scan mode not allowed on this
card
The selected scanning mode is not allowed with this module or you
have misspelled the mode parameter (see SCAN:MODE command).
2011 Empty channel list No valid channels are specified in the <channel_list>.
2012 Invalid Channel Range Invalid channel(s) specified in SCAN <channel_list> command.
Attempting to begin scanning when no valid channel list is defined.
2017 Config error 17, Slot 0 functions
disabled
Attempt to run a downloaded scan list with ARM:COUNt set to a value
other than 1. Applies to FET switches only.
2600 Function not supported on this
card
Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.
2601 Channel list required Sending a command requiring a channel list without the channel list.