Advantech RS-485 Barcode Reader User Manual


 
6-52 ADAM-5000
Command Set
$aaSi1 $aaSi1
Name Zero Calibration
Description Calibrates a specified analog input module to correct for
offset errors
Syntax $aaSi1(cr)
$ is a delimiter character.
aa (range 00-FF) represents the 2-character hexadecimal
address of the ADAM-5000 system which is to be
calibrated.
Si identifies the I/O slot which is to be calibrated.
1 represents the zero calibration command.
(cr) is the terminating character, carriage return (0Dh)
Response !aa(cr) if the command is valid.
?aa(cr) if an invalid operation was entered.
There is no response if the module detects a syntax error
or communication error or if the specified address does
not exist.
! delimiter character indicating a valid command was
received.
? delimiter character indicating the command was
invalid.
aa (range 00-FF) represents the 2-character hexadecimal
address of an ADAM-5000 system.
(cr) is the terminating character, carriage return (0Dh)
Note: In order to successfully calibrate an analog input
module's input range, a proper calibration input signal
should be connected to the analog input module
before and during the calibration process. (See also
Chapter 4, Section 4.5 on Calibration)
5017/5018 Analog Input