Gilson 215 Robotics User Manual


 
2-10
Installation 2
Plumbing Connections
Installing the rinse station
The base of one rinse station is shipped already
secured to the locator plate and is located at the
rear of the locator plate. The locator plate can hold
an optional second rinse station or you can move
the rinse station base to the alternate location.
Before installing the rinse station, make sure the
locator plate has been properly installed with the
previously-installed rinse station base located at the
rear of the instrument bed.
To install the rinse station onto the base, follow
these steps:
1 Align the triangle on the bottom of the rinse
station with the base.
2 With the rinse station’s fittings facing you,
insert the rinse station into the base.
3 Press down and turn the rinse station clockwise.
The rinse station is secure when you feel the
rinse station snap into place. When installed
correctly, the fittings point toward the control
cabinet.
4 If you will be doing shallow- or deep-pocket
rinses, connect waste tubing to the barbed
fitting installed on the rinse station.
If you will be doing flow-through rinses,
remove the plug installed on the rinse station
and replace it with a barbed fitting. Connect
tubing between the barbed fitting and the
external liquid source.