Gilson 215 Robotics User Manual


 
2-22
Installation 2
Baud rate/mode
As a default, the baud rate/mode is set to 6,
indicating that the liquid handler is set for a baud
rate of 19200 and is a master device.
You’ll need to change the selection to 0 (zero) if the
liquid handler is connected via the GSIOC to a
Gilson system and is being controlled by the 506C
System Interface from Gilson control software. A
setting of 0 indicates the liquid handler is a slave
device and the baud rate is being clocked
externally.
Note that other selections are available if the liquid
handler is being controlled by non-Gilson
applications. Refer to the table shown on the liquid
handler’s rear panel; see page 2-16.
To change the baud rate/mode:
1 Gently insert a small flat blade screwdriver into
the SW2 selector on the rear panel and turn it.
2 Align the white dot with one of the indicated
numbers.
Electrical Connections