Gilson 215 Robotics User Manual


 
C-1
C
GSIOC Commands
The Gilson Serial Input Output Channel (GSIOC) is an asynchronous serial communications
interface that enhances the power of your Gilson equipment.
The GSIOC incorporates an EIA RS-485 interface and allows up to 32 slave devices to be
controlled from a single master in a multi-drop configuration.
Each slave device is identified by a unique number which must be known to the device
and to the computer. The default ID code of the Quad-Z 215 is 22.
To control the Quad-Z 215 Liquid Handler via the GSIOC interface, you will need the
following:
a personal computer with Microsoft
®
Windows
®
98 or Windows NT
®
and any Gilson
control software or 706 Device Driver Software installed
an unused RS-232 communication port
From the controller, you:
specify the Quad-Z 215 as the device you want to control
issue commands that set operating parameters, control operation, or request
information from that instrument.