4-10
Maintenance 4
Installing different type of probe
To install a replacement probe of a different type
than is currently installed, you may want to obtain
a new probe guide insert for precise XY probe
accuracy.
1 Remove the transfer tubing’s 1/4"-28 fitting
connected to the top of the isolation probe
holder.
2 Grasp the current probe and push it up through
the top of the isolation probe holder.
3 Remove the current probe guide insert by
turning it counterclockwise. Then place the
new probe guide insert into probe guide and
secure it by turning it clockwise.
4 Insert the new probe into the top of the
isolation probe holder and pull it through the
holder and the new probe guide insert until the
tip of the probe is in the probe guide.
5 Replace and tighten the 1/4"-28 fitting.
Replacing Parts