Gilson 215 Robotics User Manual


 
3-8
Operation 3
Configuring the Liquid Handler
Rinse Site
The instrument stores the location of a rinse site in
memory. This allows the instrument to move to this
location before homing the syringe and prevents
the spilling of waste liquid or rinse diluent.
You must modify the rinse site coordinates if you
have a Quad-Z 215 Liquid Handler. Refer to the
following table to select the correct X, Y, and Z
coordinates depending on the kind of probe rinse
that will be performed. Coordinates for the rinse
site are provided for the X, Y, and Z axis in
millimeters. The coordinates identify the rinse site
for the left-most probe installed on the Quad-Z 215
Liquid Handler.
XYZ
Shallow-pocket rinse 82.9 16.5 100.0
Deep-pocket rinse 82.9 3.8 47.5
Flow-through rinse 82.9 3.8 47.5