Troubleshooting 5
Liquid Level Detector
Liquid Level Detector
Liquid level detector not detecting liquid level
Ensure that the level sensing cables are plugged
in.
Check sensitivity setting in the 215 Setup Utility
(see page 3-9) and lower the percentage.
Check if liquid is detectable. Liquid level detection
works only if there is electrical conductivity in
your liquid. Liquid level detecting will not work
with most non-polar liquids. For intermediate
polarity liquids and polar liquids, check the
sensitivity setting in the 215 Setup Utility.
Call Gilson if this is caused by faulty circuitry.
Liquid level detector falsely detecting liquid
level
Ensure that probes are installed correctly.
Check sensitivity setting in the 215 Setup Utility
(see page 3-9) and raise the percentage.
Call Gilson if this is caused by faulty circuitry.