SRS Labs SR850 Computer Hardware User Manual


 
5-27
Aux Outputs Menu
A log sweep from a higher voltage to a lower voltage is just the reverse
progression as the sweep from lower to higher. More time will be spent at
lower voltages.
Start If the selected aux output is in a sweep mode, then pressing this key will
select the start voltage as the active entry field. The value may be adjust-
ed with the knob or entered with the keypad. The start voltage can range
from 0.001 V to 21 V.
Stop If the selected aux output is in a sweep mode, then pressing this key will
select the stop voltage as the active entry field. The value may be adjust-
ed with the knob or entered with the keypad. The stop voltage can range
from 0.001 V to 21 V.
Offset If the selected aux output is in a sweep mode, then pressing this key will
select the offset voltage as the active entry field. The value may be
adjusted with the knob or entered with the keypad. The offset voltage can
range from -10.5 V to 10.5 V.
Trigger Starts? This key selects the scan start mode. Choosing Yes allows the rear
panel trigger input to start a data scan and frequency or aux output
sweep. If the sample rate is internally set, then samples are taken at the
sample rate and further triggers are ignored. If the sample rate is trig-
gered, then the first trigger starts the scan and data is sampled at every
subsequent trigger. Swept parameters are also changed at every trigger.
Choosing No requires that scans be started with the [START/CONT] key
or via an interface command.
Sweep Limits
Start:
1.000 V
Stop:
2.000 V
Offset:
0.000 V