7-14
Performance Tests
3) For each of the amplitudes listed below, perform steps 3a through 3c.
AC Calibrator Amplitudes R Output Expand
1.0000 Vrms 1
100.00 mVrms 10
10.000 mVrms 100
4.000 mVrms 250
a) Set the AC calibrator to the amplitude in the table.
b) Press the keys in the following sequence:
<Expand>
Highlight the R expand.
Enter the R Output Expand from the table.
c) Wait for the R reading to stabilize. Record the value of R.
4) This completes the amplitude linearity test. Enter the results of this test in the test record at the end of
this section.