SR850 Performance Test Record
5. Amplitude Linearity
Sensitivity Calibrator Ampl. Lower Limit Reading Upper Limit
1 V 1.0000 Vrms 0.9900 V _______ 1.0100 V
100.00 mVrms 0.0990 V _______ 0.1010 V
10.000 mVrms 0.0098 V _______ 0.0102 V
4.000 mVrms 0.0038 V _______ 0.0042 V
6. Frequency Accuracy
Input Frequency Lower Limit Reading Upper Limit
10 kHz 9.990 kHz _______ 10.010 kHz
7. Phase Accuracy
Frequency Lower Limit Reading Upper Limit
10 Hz -1.0 deg _______ +1.0 deg
100 Hz -1.0 deg _______ +1.0 deg
1 kHz -1.0 deg _______ +1.0 deg
10 kHz -1.0 deg _______ +1.0 deg
8. Sine Output Amplitude and Flatness
Sensitivity Sine Output Ampl. Lower Limit Reading Upper Limit
1 V 1.000 Vrms 0.9800 V _______ 1.0200 V
200 mV 0.200 Vrms 196.00 mV _______ 204.00 mV
50 mV 0.050 Vrms 49.000 mV _______ 51.000 mV
10 mV 0.010 Vrms 9.800 mV _______ 10.200 mV
Sine Ampl. Frequency Lower Limit Reading Upper Limit
1.000 Vrms 24 kHz 0.9800 V _______ 1.0200 V
48 kHz 0.9800 V _______ 1.0200 V
72 kHz 0.9800 V _______ 1.0200 V
96 kHz 0.9800 V _______ 1.0200 V
9. DC Outputs and Inputs
Output Offset Lower Limit Reading Upper Limit
CH1 -100.00 9.960 V _______ 10.040 V
-50.00 4.960 V _______ 5.040 V
0.00 -0.020 V _______ 0.020 V
50.00 -5.040 V _______ -4.960 V
100.00 -10.040 V _______ -9.960 V
Page 2 of 4
7-26