6-32
Remote Programming
LIA STATUS BYTE bit name usage
0 INPUT/RESRV Set when an INPUT or RESRV overload is
detected.
1 FILTR Set when a FILTR overload is detected.
2 OUTPT Set when an OUTPT overload is detected.
3 UNLK Set when a reference unlock is detected.
4 RANGE Set when the detection frequency switches
ranges (harmonic x ref. frequency decreases
below 199.21 Hz or increases above
203.12 Hz). Time constants above 30 s and
Synchronous filtering are turned off in the upper
frequency range.
5 TC Set when the time constant is changed indirect-
ly, either by changing frequency range, dynamic
reserve or filter slope.
6 TRIG Set when unit is triggered. Only if samples or
scans are triggered.
7 PLOT Set when a plot is completed.
The LIA Status bits stay set until cleared by reading or by the ❊CLS command.
ERROR STATUS BYTE bit name usage
0 Prn/Plt Error Set when an error occurs during printing or
plotting.
1 Backup Error Set at power up when the battery backup has
failed.
2 RAM Error Set when the RAM Memory test finds an error.
3 Disk Error Set when an error occurs during a disk or file
operation.
4 ROM Error Set when the ROM Memory test finds an error.
5 GPIB Error Set when GPIB fast data transfer mode aborted.
6 DSP Error Set when the DSP test finds an error.
7 Math Error Set when an internal math error occurs.
The Error Status bits stay set until cleared by reading or by the ❊CLS command.