DHZS (?) i {, j} The DHZS command sets or queries the display horizontal scale. The
parameter i selects the Full (i=0), Top (i=1) or Bottom (i=2) display and is
required. The parameter j selects a scale listed below. An error is gener-
ated if the DHZS command tries to set the horizontal scale of a display
which is not on the screen. The horizontal scale only affects chart display
types.
i time/div i time/div
0 2 mS 17 200 S
1 5 mS 18 5 min
2 10 mS 19 500 S
3 20 mS 20 10 min
4 50 mS 21 1 kS
5 0.1 S 22 20 min
6 0.2 S 23 2 kS
7 0.5 S 24 1 hour
8 1.0 S 25 5 kS
9 2.0 S 26 2 hour
10 5.0 S 27 10 kS
11 10 S 28 3 hour
12 20 S 29 20 kS
13 50 S 30 50 kS
14 1 min 31 100 kS
15 100 S 32 200 kS
16 2 min
The minimum scale is related to the sample rate. The minimum scales is
(1/sample rate) per division. This displays a minimum of 10 points on the
chart. The maximum scale is also related to the sample rate. The scale
cannot exceed that which would display the entire buffer on the chart at
once.
RBIN? i The RBIN?i command queries the bin number at the right edge of the
Full (i=0), Top (i=1) or Bottom (i=2) chart display. The selected display
must be a chart display. RBIN? along with CBIN can be used to position
the time window of the active chart display over a specific trace region.