RUNTime:INITiate:IMM
Initiates the run-time sequence and control of a measurement. All
parameters must be set up prior to initiating
RUNT:INIT:IMM. Once the
measurement is initiated, it will run until completion or until a timeout has
occurred. If a timeout does occur, it will report the error to the error stack.
Default The event trigger is defaulted to TTL. See other related commands for their
defaults.
Syntax RUNTime:INITiate:IMM;
Example RUNT:INIT:IMM;
RUNTime:SOURce:COUNt
Set the number of frequency points to be measured during the run-time
sequence. For each frequency point, each entry in the scan list will be
asserted and a trigger issued. If the SOURCE triggering method is IMM, no
source TTL trigger signal is issued and the 8530A is assumed to control the
sources. If the source triggering is TTL, a TTL signal is issued and the
run-time sequence waits for the source ready line.
Default The default is 1.
Syntax RUNTime:SOURce:COUNt <number of frequency points>;
Example RUNT:SOUR:COUN 1;
RUNT:SOUR:COUN 51;
NOTE For CW angle scan measurements the number of frequencies should be set
to 1.