Agilent Technologies 85330A Network Card User Manual


 
2-2 85330A Multiple Channel Controller
Performance Verification
Recommended
Equipment
Table 2-1 lists the equipment that is mandatory when performing the
performance verification.
Table 2-1 Required Equipment
Qty Item Agilent Part or Model Number
1
1 Downloadable Driver Disk, rev A.03.00 85330-10016
2
1 Multimeter (20 Vdc, digital) E2377A
1 Digital Oscilloscope (100 MHz bandwidth
1 MHz Single-shot bandwidth)
54501A
1 Tee adapter (BNC male to female, female) 1250-0781
3 Cable (BNC male to male 122 cm) 8120-1840
1
Network Analyzer System
3
(50 MHz to 40 GHz
4, 5
90 dB dynamic range)
8722C (Option 003)
1
Plotter or Printer
6
HP 7550A+, or LaserJet
1 Torque wrench (2.4/3.5 mm @ 8 in-lb)
8710-1765
7
1. Other part numbers or model numbers may be used if they meet the minimum requirements.
2. Included with 85330A shipment.
3. System must include vector network analyzer, test port cables, and a calibration kit.
4. A 50 MHz to 26.5 GHz network analyzer may be used if the 3.5 mm adapters are used with the
microwave switches.
5. If a network analyzer is used with a frequency range less than the switches, the performance
verification will only be valid over the frequency range of the network analyzer.
6. Plotter or printer must be connected to the network analyzer and oscilloscope for test documentation.
7. Use this wrench for any 2.4/3.5 /SMA mm connections in this procedure.