Agilent Technologies 85330A Network Card User Manual


 
85330A Multiple Channel Controller 7-13
Remote Programming Command Reference
Standard Command Reference
RUNTime:SWITch:SCAN
Set up the list of channel addresses for run-time control. For each event
increment signal received by the 85330A, each channel address entry will be
asserted and a subsequent trigger issued to begin a measurement of each data
point.
This command can also be used to set the TTL level of the rear panel AUX 2
connector during run-time mode.
Default Scan list for port 1 and port 2 are empty.
AUX 2 connector default is TTL LOW.
Syntax RUNT:SWIT:SCAN 1 | 2,(@ <switch address list>);
or
RUNT:SWIT:SCAN 3,(@0 | 1);
This command controls port AUX 2.0 sets the AUX 2 BNC to a TTL low.
A1 sets AUX2 to a TTL high.
NOTE AUX 1 cannot be controlled in RUNTime mode.
Example RUNT:SWIT:SCAN 1,(@101,102,101,102);
RUNT:SWIT:SCAN 2,(@101,101,102,102);
RUNT:SWIT:SCAN 3,(@0);
Sets AUX 2 output to TTL LOW during run-time.
NOTE Switch address list requirements are the same as the ROUT:CLOS
command. A switch address scan list can remain empty and no data will be
written to the port. If a port is not empty, then an error will be issued if the
scan lists for both ports do not contain the same number of entries.