Agilent Technologies 85330A Network Card User Manual


 
85330A Multiple Channel Controller 2-1
Performance Verification
2 Performance Verification
In This Chapter Recommended performance verification test equipment
Performance verification tests
Verifying the multiple channel controller
Verifying the RF switches
Description This performance verification should be done once a year, or more
frequently as needed. This procedure verifies that the Agilent 85330A
system meets or exceeds its published specifications. Specifications are
listed in Chapter 4, “General Information.” If traceability to a local standards
organization is required, then it can be done with this procedure. One such
standards organization is the United States National Institute of Standards
and Technology (NIST, formerly NBS).
This verification can be performed at the measurement system’s site, or at a
service bench. The verification will verify all parts of the multiple channel
system: multiple channel controller, switches, and cables. This verification
does not check the performance of the measurement system’s transmitter or
receiver. Use the measurement system’s manual for more information on
that procedure.