Kodak RP 30 Printer User Manual


 
Operator’s Guide
Monthly Maintenance
Checking the Solution Temperatures
1. Touch:
Settings
Machine settings
PP temperature
2. Set the reference temperatures for the solutions and the dryer, if necessary:
DEV 40 °C ± 0.3
BX 38 °C ± 2
STB 37 °C ± 2
Dryer 65 °C
3. Calibrate the temperature sensors:
a. Touch Readjust xxx.
A window with a numeric keypad opens. The temperature of the selected
solution is indicated in the first line of the input field.
b. Remove the chemical filter.
Using a calibrated thermometer, measure the current temperature of the
processing solution through the free opening of the filter holder.
c. Enter the current temperature value via the numeric keypad.
The sensor will be calibrated after pressing OK.
NOTE: The Reset reference and Reset calibration buttons are used to enter the
standard values of the database.
8-22 KODAK PROFESSIONAL RP 30 Laser Printer