Acquiring and Displaying Waveforms
TDS 500C, TDS 600B, & TDS 700C User Manual
3–9
Do the following steps to automatically set up the oscilloscope:
1. Press the channel selection button (such as CH 1) corresponding to your
input channel to make it active.
2. Press AUTOSET.
If you use Autoset when one or more channels are displayed, the oscilloscope
selects the lowest numbered channel for horizontal scaling and triggering.
Vertically, all channels in use are individually scaled. If you use Autoset when no
channels are displayed, the oscilloscope will turn on channel one (CH 1) and
scale it.
NOTE. Autoset may change vertical position in order to position the waveform
appropriately. It always sets vertical offset to 0 V.
If a standard mask is active, Autoset adjusts the selected trace to match the
mask, if possible. Vertical scale and offset, horizontal scale, trigger position, full
bandwidth, average, and trigger parameters are set as required for the standard.
If a calibrated optical probe is attached to Ch1 and an OC or FC standard is
selected, Ch 1 is selected and other channels are turned off.
Table 3–1 lists the autoset defaults.
Table 3–1: Autoset defaults
Control Changed by autoset to
Selected channel Numerically lowest of the displayed channels
Acquire Mode Sample
Acquire Repetitive Signal
(TDS 500C and 700C Models
Only)
On
Acquire Stop After RUN/STOP button only
Deskew, Channel/Probe Unchanged
To Autoset the
Oscilloscope
List of Autoset Defaults