Troubleshooting
Troubleshooting Gateway Configuration
4 Verify the BOOTP server is on the same local subnet as the
Gateway. If not, consult your Network Administrator to verify that
your network configuration supports using a BOOTP server that is
not on the same subnet.
5 When the Gateway is powered on, verify there is a fast flash on the
LAN Conn LED.
Troubleshooting BOOTP with TFTP Configuration
If you have trouble configuring the LAN/GPIB Gateway using the BOOTP
with TFTP configuration method:
1 Follow the steps in “Troubleshooting BOOTP Configuration”.
2 Verify that the TFTP server system is configured to run the tftpd
daemon. See the Administering ARPA Services manual for more
information.
3 Consult your BOOTP server documentation to verify you have the
correct permissions on the specified TFTP configuration file.
4 Verify that the Gateway’s configuration values are correct in the
TFTP file.
5 Look for errors specific to the TFTP values in the syslog file on the
configured syslog server or with the Telnet syslog-display
command.
NOTE
Your BOOTP server may support the bootpquery capability to verify
your BOOTP configuration. See your BOOTP server documentation for