STATus Subsystem
3–54
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:PRESet
Presets the SCPI enable registers. The Operation and Questionable Enable
registers are preset to zero, disabling all events. All other enable registers are
preset to all ones, enabling all events. The Error/Event Queue is preset to the
default value of –499:0. The SCPI Positive Transition Filters are preset to 1.
This command does not clear the SCPI event registers or the IEEE 488.2 enable
and event registers.
STATus:PRESet
0
None
None.
Command: STAT:PRES
STATus:OPERation:ENABle
STATus:QUEue:ENABle
STATus:QUEStionable:ENABle
*CLS
Syntax
Reset Value
Errors and Events
Dependencies
Examples
Related Commands