STATus Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–93
STATus:QUEStionable:VOLTage:CONDition?
Queries the contents of the Voltage Condition register. A value of one in a bit
indicates the corresponding power supply has a failure. Unused bits always
return a value of zero. The definition of the bits in this register can be found in
STATus:QUEStionable:VOLTage? command on page 3–91.
Reading the register does not clear it.
There is no command associated with this query.
STATus:QUEStionable:VOLTage:CONDition?
Parameter Query response
Not Applicable <NR1>
0
None
None
Query: STAT:QUES:VOLT:COND?
Response: 0
STATus:QUEStionable:VOLTage?
STATus:QUEStionable:VOLTage:ENABle
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples