STATus Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–33
STATus:OPERation?
Returns the contents of the Operation Event register. A value of one in a bit
indicates the corresponding condition bit changed from false to true. Unused bits
always return a value of zero. The register is cleared by the *CLS command.
There is no command associated with this query.
STATus:OPERation[:EVENt]?
Parameter Query response
Not applicable <NR1>
Operation status word bits Definition
D0 CALibrating
D1 Not Used
D2 Not Used
D3 Not Used
D4 Not Used
D5 VXI Triggers
D6 Not Used
D7 Not Used
D8 VXI Monitor
D9 Mainframe Power Down
D10 VXI Interrupts
D11 Not Used
D12 Not Used
D13 Not Used
D14 Not Used
D15 Not Used
0
None
Reading the register clears it.
Syntax
Parameters
Reset Value
Errors and Events
Dependencies