VXI VX1420A Computer Hardware User Manual


 
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
4–1
Specifications
This chapter contains the complete specifications for the IntelliFrame Main-
frame. Within each section, the specifications are arranged in functional groups
such as: AC Power Source, Secondary Power, Cooling, Certifications and
compliances, Environmental, Backplane, Display System and Mechanical.
All specifications are warranted unless unless labeled typical. Typical specifica-
tions are provided for your convenience but are not guaranteed. Specifications
that are marked with the n symbol are checked in the Performance Verification
chapter beginning on page 5–1.
NOTE. All VX1410A specifications listed in this chapter also apply to VX1411A
mainframes that have been upgraded. All VX1420A specifications also apply to
upgraded VX1410A and VX1411A mainframes.
Table 4–1: AC Power Source
Characteristic Description
Source Voltage 90 V
RMS
to 250 V
RMS
, continuous range
Source Frequency 115 V Operation: 45 to 66 Hz or 360 to 440 Hz
230 V Operation: 45 to 66 Hz
Power Consumption VX1410A:1350 W line power maximum
VX1420A:1450 W line power maximum
Fuse Rating
90 V – 132 V Operation 0.25 in × 1.25 in, Slow Blow, 20 A, 250 V
103 V – 250 V Operation 0.25 in × 1.25 in, Fast Blow, 15 A, 250 V
207 V – 250 V Operation 5 mm × 20 mm, Fast Blow, 6.3 A, 250 V
Inrush Surge Current 70 A maximum
Input Current VX1410A:15 A maximum at 90 V
RMS
6.3 A maximum at 207 V
RMS
VX1420A:16.5 A maximum at 90 V
RMS
6.3 A maximum at 207 V
RMS
Power Factor Correction 0.99 @ 60 Hz operation
0.95 @ 400 Hz operation
Power Disconnect Front Panel On/Standby. No primary switch on rear
panel. Power cord provides main power disconnect. The
Front Panel On/Standby switch may be disabled for
remote operation. Enhanced monitor provides ability to
turn the instrument on/off under program control.