Table 4–4: Certifications and compliances
Characteristic Description
EC Declaration of Conformity Meets intent of Directive 89/336/EEC for Electromagnetic
Compatibility and Low Voltage Directive 73/23/ECC for Product
Safety. Compliance was demonstrated to the following
specifications as listed in the Official Journal of the European
Communities:
EMC Directive 89/336/EEC:
EN 55011 Class A Limits for Conductive Emissions.
EN60555-2 AC Power Line Harmonic Emissions.
EN 55011 Class A Limits for Radiated Emissions.
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge Immunity
IEC 801-3 RF Electromagnetic Field Immunity
IEC 801-4 Electrical Fast Transient/Burst Immunity
IEC 801-5 Power Line Surge Immunity
Low Voltage Directive 73/23/EEC:
EN61010-1 Safety requirements for electrical equipment
for measurement, control, and laboratory use
Australian Declaration of
Conformity (C-Tick Mark)
Complies with Australian EMC Framework, under the Standard
applicable product class 2064, governing radiated and
conducted emissions.
Immunity, Enclosure,
Electrostatic Discharge
(ESD)
Up to 8 kV with no change to control settings or impairment of
normal operation, and up to 15 kV with no loss of stored data or
damage that prevents recovery of normal operation by the user
(IEC 801-2).
Immunity, Enclosure, Radio
Frequency Electromagnetic
Field
No mainframe failures when the mainframe is subjected to a
3 V/m electromagnetic field over the frequency range of 27 MHz
to 500 MHz (IEC 801-3).
Immunity, Fast Transients,
Common Mode
No loss of stored data, change to control settings, degradation
of performance, or temporary loss of function will occur when the
mainframe is subjected to the following transients (IEC 801-4):
Port Peak Voltage Tr/Th Rep Frequency
Signal & Control 0.5 kV 5/50 ns 5 kHz
DC Power 0.5 kV 5/50 ns 5 kHz
AC Power 1.0 kV 5/50 ns 5 kHz