VXI VX1420A Computer Hardware User Manual


 
Status and Events
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–177
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
POWer Status
Not used
STATus:OPERation:POWer:[:EVENt]?
STATus:OPERation:POWer:CONDition?
STATus:OPERation:POWer:ENABle
Summary of POWer
Total of all supplies
Figure 3–15: Power status register
The Current Status register provides status information on the individual power
supply Amperages (see Figure 3–16). You can detect problems with any power
supply current by monitoring the corresponding bit in the Current Status register.
If a problem occurs, the corresponding bit will be set in the register.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CURRent Status
Not used
STATus:OPERation:CURRent:[:EVENt]?
STATus:OPERation:CURRent:CONDition?
STATus:OPERation:CURRent:ENABle
Summary of CURRent
+24V Supply
+12V Supply
+5V Supply
–2V Supply
–5.2V Supply
–12V Supply
–24V Supply
Figure 3–16: Current Status register
The Voltage Status register provides status information on the individual power
supply voltages (see Figure 3–17). You can detect problems with any one of the
power supply voltages by monitoring the corresponding bit in the Voltage Status
register. If a problem occurs, the corresponding bit will be set in the register.