SYSTem Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–143
SYSTem:SYSReset
SYSTem:SYSReset?
Sets or queries whether the VXI SYSRESET signal can reset the Enhanced
Monitor.
When SYSRESET is enabled, asserting the VXI SYSRESET signal will reset
the Enhanced Monitor. When disabled, asserting the VXI SYSRESET signal
only initializes the Enhanced Monitor VXI Interface; all other Enhanced Monitor
functions are not affected.
The factory default setting is disabled (0).
NOTE. If SYSRESET is enabled and the mainframe is powered down, commu-
nication through the RS-232 port is not possible even if an external +5 V is
supplied to the enhanced monitor card. The display will not show the current
status and the scroll buttons will be inoperative. Consequently the enhanced
monitor is inoperative.
SYSTem:SYSReset <Boolean>
SYSTem:SYSReset?
<Boolean> Query response
<NRf>
N ≠ 0
0
ON
OFF
<NR1>
1
0
1
0
Not Applicable
–224, “Illegal Parameter Value”
You attempted to set the parameter to an illegal value.
None
Command: SYST:SYSR ON
Query: SYST:SYSR?
Response: 1
SYSTem:SYSReset:IMMediate
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands