STATus Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–39
STATus:OPERation:VXIInterrupt:CONDition?
Queries the contents of the VXI Interrupt Monitor Condition register. A value of
one in a bit indicates the corresponding condition bit has transitioned from false
to true. Unused bits always return a value of zero. Definition of the bits in this
register can be found in STATus:OPERation:VXIInterrupt command.
The enhanced monitor latches the state of the VXI IRQ lines into a hardware
register. This register is cleared upon VXI Monitor initialization. This query
returns the state of this register.
There is no command associated with this query.
STATus:OPERation:VXIInterrupt:CONDition?
Parameter Query response
Not applicable <NR1>
Not Applicable
None
None
Query: STAT:OPER:VXII:COND?
Response: 0
STATus:OPERation:VXIInterrupt?
STATus:OPERation:VXIInterrupt:ENABle
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands