VXI VX1420A Computer Hardware User Manual


 
STATus Subsystem
3–86
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:QUEStionable:TEMPerature:CONDition?
Queries the contents of the Temperature Condition register. A value of one in a
bit indicates the corresponding temperature condition is outside of the limits.
Unused bits always return a value of zero. Definition of the bits in this register
can be found under the STATus:QUEStionable:TEMPerature? command on
page 3–83.
If the fan speed is already set to the maximum speed and the delta temperature is
above the limit for 30 seconds, a failure will be reported.
Reading the register does not clear it.
There is no command associated with this query.
STATus:QUEStionable:TEMPerature:CONDition?
Parameter Query response
Not Applicable <NR1>
0
None
None
Query: STAT:QUES:TEMP:COND?
Response: 0
STATus:QUEStionable:TEMPerature?
STATus:QUEStionable:TEMPerature:ENABle
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples