VXI VX1420A Computer Hardware User Manual


 
STATus Subsystem
3–106
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:SEVent?
Returns the summary information on the power supply voltage events, tempera-
ture events, blower events, filter events, and VXI monitor events.
A one in a bit indicates the corresponding event has been asserted. Unused bits
always return a value of zero. The query returns a value of zero at power on or
when you send the *CLS command. Reading events by this command does not
clear the events. Reading the specific event register will clear the corresponding
bits in the register. (That is, STAT:QUES:VOLT? will clear all of the voltage bits
in this register). Summary information for current, power, triggers and IRQ
reside in an overflow register. If bit 30 is set, the overflow register should be
queried for more information. See the STATus:S2EVent? query on page 3–108.
There is no command associated with this query.
STATus:SEVent?
Parameter Query response
Not applicable <NR1>
Event status word bits Definition
D0 Blower 1
D1 Blower 2
D2 Filter
D3 Slot 0 Delta Temperature
D4 Slot 1 Delta Temperature
D5 Slot 2 Delta Temperature
D6 Slot 3 Delta Temperature
D7 Slot 4 Delta Temperature
D8 Slot 5 Delta Temperature
D9 Slot 6 Delta Temperature
D10 Slot 7 Delta Temperature
D11 Slot 8 Delta Temperature
D12 Slot 9 Delta Temperature
D13 Slot 10 Delta Temperature
D14 Slot 11 Delta Temperature
D15 Slot 12 Delta Temperature
Syntax
Parameters