Input Calibration
4/07 UDC2500 Universal Digital Controller Product Manual 139
Table 5-2 Voltage and Milliamp Equivalents for Input 2 Range Values
Range Values Sensor Type PV Input Range
0 % 100 %
Linear
Milliamps
4 to 20 mA
0 to 20 mA
4.00 mA
0.00 mA
20.00 mA
20.00 mA
Volts
1 to 5 Volts
0 to 5 Volts
0 to 2 Volts
1.00 Volts
0.00 Volts
0.00 Volts
5.00 Volts
5.00 Volts
2.00 Volts
5.3 Preliminary Information
Disconnect the Field Wiring
Tag and disconnect any field wiring connected to the input (#1 or #2) terminals on the
rear of the controller.
XXXX
R
+
_
Input 1
Connections
25 R
26 +
27 –
mA+
Volt+
Input 2
Connections
22 mA+
23 V+
24 –
_
Input 1
Input 2
Figure 5-1 Input 1 and Input 2 Wiring Terminals
Equipment Needed
Table 5-3 lists the equipment you will need to calibrate the specific types of inputs that
are listed in the table. You will need a screwdriver to connect these devices to your
controller.
Table 5-3 Equipment Needed
Type of Input Equipment Needed
Thermocouple
Inputs (Ice Bath)
A calibrating device with at least ± 0.02 % accuracy for use as a
signal source such as a millivolt source.
Thermocouple extension wire that corresponds with the type of
thermocouple that will be used with the controller input.
Two insulated copper leads for connecting the thermocouple
extension wire from the ice baths to the mV source.
Two containers of crushed ice.
Thermocouple
Inputs (T/C Source)
A calibrating device with at least ± 0.02 % accuracy for use as a
signal source such as a millivolt source.
Thermocouple extension wire that corresponds with the type of
thermocouple that will be used with controller input.