NCR 7870 Scanner User Manual


 
Chapter 5: Calibration 5-11
Verifying Calibration
This section describes how to verify that the 7870 has been properly
calibrated.
No error codes should appear during calibration. The codes are shown
in the table below. If a code is displayed, the calibration is invalid and
a corrective action must be undertaken before attempting recalibration.
Error
Code
Problem Suspect Component
1 - - - - Calibration error
A/D output below 000h at digital to
analog converter
Scale Board
Load Cell
2 - - - - Calibration error
A/D output below 280h at digital to
analog converter
Scale Board
Load Cell
3 - - - - Calibration error
Analog to digital converter output
too high at 15 kg (30.0 lb.)
Scale Board
Load Cell
4 - - - - Vibration to scale during calibration Motors in checkstand
9 - - - - Interference detected during
calibration
Top Plate is touching
checkstand, scale was
jarred, or overload screw
is misadjusted.
If no error codes occur during calibration, four test procedures must be
performed to verify calibration. These procedures are, in order,
Increasing Load Test ,
Over-Capacity Test,
Decreasing Load Test, and
Shift Test.