Xantrex Technology XTR 850 Watt Power Supply User Manual


 
Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-26 975-0200-01-01
Resistive-Controlled Voltage ISOL Setup
Activating ISOL Programming Voltage Mode
To activate ISOL programming voltage mode using an external
resistor:
1. Turn the mode control knob to the VAP position or press the Rotary
knob/Enter button if the control knob is already at the VAP position.
VAPr is displayed on the output voltage display.
2. Turn the Rotary knob/Enter button to select R IS (resistor source
from isolated connector).
3. Press the Rotary knob/Enter button to commit the selected mode.
LE U is displayed on the output voltage display.
4. Select the input source level using the Rotary knob/Enter button.
5. Once the desired level has been selected, press the Rotary knob/
Enter button to commit the change.
The display will return to the normal display.
Deactivating ISOL Voltage Mode
To deactivate the ISOL voltage mode:
1. Turn the mode control knob to the VAP position or press the Rotary
knob/Enter button if the control knob is already at the VAP position.
2. UAPr is displayed on the output voltage display.
3. Turn the Rotary knob/Enter button to select nonE.
4. Press the Rotary knob/Enter button to commit the change.
The display will return to the normal display.
Query for ISOL Voltage Input Level
To query for ISOL voltage input level from non-isolated input:
1. Turn the mode control knob to the VAP position or press the Rotary
knob/Enter button if the control knob is already at the VAP position.
UAPr is displayed on the output voltage display.