Calibration and Troubleshooting
6-10 975-0200-01-01
5. Set the input source as Voltage Non-Isolated for Voltage
Programming. The SCPI command is:
SYST:REM:SOUR AVOL
6. Turn the main output to On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
7. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL
8. Press the terminal keys – or + to reduce or increase the main output
voltage value. Adjust the main output voltage to 100%, measured on
the external voltmeter on the output.
9. Press any key other than the + or – keys to exit from interactive
calibration mode.
Non-isolated Resistive Programming of Voltage Calibration
To calibrate the non-isolated resistive programming of voltage:
1. Disconnect the load and connect the voltmeter to the output.
2. Connect the 4.000kΩ resistor to the non-isolated connector, resistive
programming of voltage, and voltage programming (J1.9-J1-11, and
J1.7).
3. Set the resistive APG level to 4kΩ. The SCPI command is:
SYST:COMM:APR:LEV 4
4. Set the input source as Resistive Non-Isolated for Voltage
Programming. The SCPI command is:
SYST:REM:SOUR ARES
5. Turn the main output On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
6. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:RES
7. Press the terminal keys – or + to reduce or increase main output
voltage value. Adjust the main output voltage to 100%.