Xantrex Technology XTR 850 Watt Power Supply User Manual


 
Calibration and Troubleshooting
975-0200-01-01 6-5
6
Calibrating the Output Voltage
Gain calibration of the power supply has the greatest affect on the
accuracy in the high voltage range. Offset calibration has the greatest
affect on accuracy of the power supply at low voltages. The same
calibration command is used for the gain and offset calibrations. The type
of calibration is determined internally by the XTR depending on if the set
point is above 10% of the maximum output voltage value or not. If it is
above 10% of the maximum voltage, the calibration is a gain calibration;
if is below 10%, it is an offset calibration.
Gain Calibration
To perform gain calibration:
1. Disconnect the power supply from the load.
2. Connect the output terminals to a precision voltmeter.
3. Turn both the power supply and the voltmeter to On.
4. Set the nominal level to 90% of nominal voltage level.
5. Turn on the output power.
6. Read the voltage value on voltmeter display.
7. Type the SCPI command with voltage value as parameter “data”.
For example, type :CAL:OUTP 153.3 if you had an XTR 150-5.6
calibrating with the voltage set point set to 150 V and the voltmeter
was reading 153.3 volts.
Offset Calibration
To perform offset calibration:
1. After performing gain calibration, set the programmed voltage 10%
from nominal.
2. Read the voltage value on the voltmeter display.
3. Type the SCPI command with voltage value as parameter “data”.
For example, you would type :CAL:OUTP153.3 if you had an
XTR 150.5.6 calibration with the voltage set point set to 150V and
the voltmeter was reading 153.3 volts.
After performing offset calibration, Xantrex recommends that you repeat
gain calibration.