Calibration and Troubleshooting
6-12 975-0200-01-01
SYST:COMM:APR:LEV:CURR 4
4. Set the input source as Resistive Non-Isolated for Current
Programming. The SCPI command is:
SYST:REM:SOUR:CURR ARES
5. Turn the main output to On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
6. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:CURR:RES
7. Press the terminal keys – or + to reduce or increase the main output
current value. Adjust the main output current to 100%.
8. Press any key other than the + or – keys to exit from interactive
calibration mode.