Remote Operation
5-34 975-0200-01-01
The operation shutdown status register describes the cause of the power
supply shutting down the output. More than one bit may be active and
multiple actions will be required to restart the unit. The protection
shutdown sub-register indicates which protection mechanisms have
caused the power supply to shutdown.
Table 5-11
OPERation Status Register
Bit Bit Weight Bit Name Description
0 1 CALibrating Indicates that the supply is in CALibration Mode.
1 2 SETTling Not implemented
2 4 RANGing Not implemented
3 8 SWEeping Not implemented
4 16 MEASuring Not implemented
5 32 Waiting for Arm Not implemented
6 64 Waiting for Trigger Not implemented
7 128 CORRecting Not implemented
8 256 Not Used Not Used
9 512 SHUTdown Summary Reflects the summary of the SHUTdown Sub-
Register.
10 1024 Local Lockout Front panel lockout
11 2048 Current Share Summary Reflects the summary of the CSHare Sub-Register.
12 4096 Constant Voltage Mode Reflects the output mode of the supply.
13 8192 Constant Current Mode Reflects the output mode of the supply.
14 16384 Program Running Indicates if an Autosequence program is currently
running.