Xantrex Technology XTR 850 Watt Power Supply User Manual


 
Remote Operation
5-60 975-0200-01-01
:PROG:STAT RUN
The output after the last command might be as follows:
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Repeating an Auto Sequence Program:
An Auto Sequence Program can be configured to repeat any number of
times or infinitely. The default at power up is to have the repeat count
default to once. The program will repeat if the repeat count can be
changed using the :PROGram[<channel>]:REPeat command. If a
program is running or pause the :PROGram[<channel>]:REPeat
command will not work.
Commands:
:PROGram[<channel>]:REPeat {?|<count>|INFinity}
Where:
<count> is the repeated count from 1 to 65 534.
INFinity is the character mnemonic for repeating without end.
For Example:
:PROG:STAR
*IDN?
*IDN?
:PROG:STOP
:PROG:REP 2
:PROG:STAT RUN
The output after the last command might be as follows:
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Xantrex, XTR 150-5.6, SN# E00123456, 1.00 Build 10,
21/11/2005
Important:
Execution of the program may be terminated at any time by
pressing the Esc key in the MS Windows Hyper Terminal window. The power
supply can be controlled during program execution.